A Guide to HPLC and LC-MS Buffer Selection [PDF]

which buffer precipitation can occur. Table 3. Solubility of Potassium Phosphate, pH 7.0, in Common HPLC. Solvents1. The

9 downloads 4 Views 4MB Size

Recommend Stories


Chiral HPLC Column Selection and Method Development Guide
Don't ruin a good today by thinking about a bad yesterday. Let it go. Anonymous

guide to selection
We may have all come on different ships, but we're in the same boat now. M.L.King

Guide to Selection
Learn to light a candle in the darkest moments of someone’s life. Be the light that helps others see; i

an alternative to lcms and dad-hplc for the analysis of complex mixtures
Happiness doesn't result from what we get, but from what we give. Ben Carson

Guide to Fume Hood Selection
Don’t grieve. Anything you lose comes round in another form. Rumi

RECRUITMENT and SELECTION GUIDE
You're not going to master the rest of your life in one day. Just relax. Master the day. Than just keep

LCMS Congregation
When you talk, you are only repeating what you already know. But if you listen, you may learn something

[PDF] A Guide to SQL
Live as if you were to die tomorrow. Learn as if you were to live forever. Mahatma Gandhi

Selection Guide
Don't count the days, make the days count. Muhammad Ali

Selection Guide
Don’t grieve. Anything you lose comes round in another form. Rumi

Idea Transcript


A Guide to HPLC and LC-MS Buffer Selection

HPLC Columns

John Dolan

Your decision has lasting effects. Choose wisely.

HPLC Columns Ultra Inert Base-Deactivated HPLC Columns

For Performance, Selectivity and Guaranteed Reproducibility

ACE



performance guarantee

If ACE does not outperform your existing column (of equivalent phase, particle size and dimensions), send in your comparative data within 60 days and keep the ACE column FREE OF CHARGE.

Using Buffers with HPLC and LC-MS

Table of Contents

Page

Why Control pH?

2

Practical Implications

4

Controlling pH

4

What about LC-MS?

6

Just Interested in Low pH?

7

Or High pH?

7

How Much?

7

Solubility Issues

10

Dilution Effects

11

Buffer Preparation

12

Precautions

13

Summary

15

References

16

www.ace-hplc.com

Using Buffers with HPLC and LC-MS When samples contain ionizable compounds, the mobile phase pH can be one of the most important variables in the control of retention in a reversed-phase HPLC (RP-HPLC) separation. However, if it is not controlled properly, pH can be a source of many problems. Since most compounds analyzed by RP-HPLC contain one or more acidic or basic functional groups, most mobile phases require pH control. For this reason, buffers are widely used. This booklet highlights some of the important aspects of mobile phase pH for the practical chromatographer. Why Control pH? Figure 1 illustrates the need for pH control when ionizable compounds are present. When an acid is more than 2 pH units above or below its pKa, it will be >99% ionized or non-ionized, respectively. Bases are ionized below their pKa and non-ionized above their pKa. The non-ionized form will be less polar (more hydrophobic), and thus more strongly retained in a reversed-phase system. Thus, at low pH, acids will be more retained (Fig. 1a), whereas bases will be more retained at high pH (Fig. 1b).

Figure 1. Retention vs. pH for a hypothetical acid (a) and base (b).

If the mobile phase pH is near the pKa, you can see that small changes in pH can make large changes in retention – not what is desired for a robust separation. This is illustrated in Figure 2a, which shows the extreme sensitivity of some compounds to very small changes in pH. Here the resolution changes by a factor of two for a change of only 0.1 pH units – this is the amount of error in pH adjustment common to many laboratories. Figure 2b represents a plot of retention vs. pH for an acid, a base and a neutral compound. At pH 5, retention is less sensitive to pH than it is at pH 3 (for the acid) or pH≥6 for the base. 2

www.ace-hplc.com

Besides the instability of retention times when the pH is near the pKa, relative peak spacing (selectivity) can change if compounds of similar structure are present.

(a)

(b)

Figure 2. The effect of small changes in mobile phase pH on the separation. (a) Basic analytes: p-anisidine, m-toluidine, 4-chloroaniline, 3-aminobenzonitrile (in retention order); 27:73 methanol/phosphate buffer. (b) Acid: salicylic acid; base: methylamphetamine; neutral: phenacetin.

Another factor that should be considered when choosing the mobile phase pH is the stability of the column. As a general rule, silica-based columns should be operated at 27, so operating at a sufficiently high pH to suppress ionization will be detrimental to most columns. So, all other things being equal, it is best to start out at a low pH. If you need to operate the column at a high pH, be sure to select a column known to be stable in the pH region you choose. If low-pH ion suppression does not provide acceptable results, the mobile phase pH can be adjusted to help obtain the desirable separation. It usually is most fruitful to adjust the mobile phase organic content (%B-solvent) to obtain acceptable retention for neutral and non-ionized compounds, then to adjust the pH to fine-tune retention of ionic analytes. Controlling pH Since the retention of ionizable compounds is very sensitive to the mobile phase pH, it is necessary to control the pH of the mobile phase by the addition of a buffer. A buffer maintains the pH when a small amount of acid or base is added. Many different substances have been used for buffering in HPLC; some of these 4

www.ace-hplc.com

additives are listed in Table 1. A buffer is most effective when used within ±1 pH unit of its pKa, but may provide adequate buffering ±2 pH units from the pKa. Table 1. pKa Values of Common Mobile Phase Additives1 pKa (25°C) 0.3 2.15 3.13 3.75 4.76 4.76 4.86 6.35 6.40 7.20 8.06 9.23 9.25 9.78 10.33 10.72 11.27 12.33

compound trifluoroacetic acid2 phosphoric acid (pK1) citric acid (pK1) formic acid acetic acid citric acid (pK2) propionic acid carbonic acid (pK1) citric acid (pK3) phosphoric acid (pK2) tris boric acid ammonia glycine (pK2) carbonic acid (pK2) triethylamine pyrrolidine3 phosphoric acid (pK3)

1

data of [1]; 2 Merck Index; 3 CRC Handbook of Chemistry and Physics

The most popular buffers for HPLC with UV detection are phosphate and acetate. Phosphate and acetate are particularly useful buffers because they can be used at wavelengths below 220 nm. As can be seen from Table 2, phosphate has three pKa values that give it three buffering ranges: 1.1

Smile Life

When life gives you a hundred reasons to cry, show life that you have a thousand reasons to smile

Get in touch

© Copyright 2015 - 2024 PDFFOX.COM - All rights reserved.