Idea Transcript
Determination of Calcium and Magnesium in Serum, Urine, Diet, and Stool by Atomic Absorption Spectrophotometry Elizabeth G. Gimblet, Amy F. Marney,
and Roy W. Bonsnes
Atomic absorption spectrophotometry was evaluated as a method for the determination of calcium and magnesium in serum, urine, diet, and stool, and was found a most suitable technic for a large clinical chemistry laboratory.
was developed by Walsh and his associates (1) as an analytical technic applicable to the determination of many metals. Willis (2-4) found that this technic was applicable in particular to tile determination of calcium and magnesium in Serum and urine. These published results indicated that atomic absorption spectrophotometry would be a simple, rapid, and practical method for tile determination of calcium and magnesium in biologic material, particularly in serum and urine, and that it could be adapted effectively to the needs of a clinical chemistry laboratory. After evaluation, we have used atomic absorption spectrophotometry for routine determinat ion of calcium and magnesium iii serum and urine since November 1962. Since our studies of this technic began, several reports have appeared on the determination of calcium and magnesium in biologic materials (5-12), indicating considerable variation in tile methods used. Thus, it seems appropriate to publish the procedures we have found useful for routine determinations of these elements, not only in serum and urine, but also in diet and stool. ATOMIC
ABSORPTION
Instrumentation The instrument the Perkin-Elmer
used in this work, and still in routine operation, is Model 214 atomic absorption spectrophotometer (13),
From the Laboratory for Clinical Cimemistry, Time New York Hospital, and the Chemistry Laboratory of the Department of Obstetrics and Gynecology, The New York Hospital and Cornell University Medical College, New York, N. Y. 10021. Some of the data on the determination of serum magnesiimmn are from a dissertation in 1962 by the senior author in partial fulfillment of the requirememmts for the degree of Master of Arts, Hunter College, New York. Received for publication Sept. 2, 1966; accepted for publication Oct. 27, 1966. 204
Vol. 13, No. 3, 1967
205
CALCIUM AND MAGNESIUM
connected to a Sargent Model SR recorder. All data recorded were collected with tile burner described by Slavin (14). A lean acetylene-air flame has beell used for all analyses. The air compressor used supplies oil-free air. Both calcium and magnesium are determined with the use of hollow cathode lamps. The 4226-A calcium line and the 2852-A magnesium line are isolated with the monochromator of the instrument set to a spectral slit width of about 300 . The settings of the instrument for the burner, the hollow cathode lamps, and the external recorder depend upon the particular components used. Some details on these components with respect to the determination of calcium are reported by Slavin et al. (15). In general, the difficulties encountered with the instrument, including the burner, have been no greater than those an experienced laboratory worker might expect with any laboratory instrument or apparatus of equal complexity. Operation of tile instrument ilas been largely troublefree. The slot of tile burner, which is 0.020 in., tends to become constricted slightly when sera are analyzed. Currently, 27 ± 9 sera are analyzed for calcium and roughly 2 ± 2 for magnesium each day the instrument is in use. Under these operating conditions and Witil tile diluents described, the burner is cleaned once a day. How many more sera could be analyzed without tile burner’s requiring cleaning remains to be determined.
Standards and Samples Calcium carbonate (Mallinckrodt’s Primary concentrated IiCl served as tile stock standard solution was diluted appropriately to obtain made up to contain the equivalent of 2.5, 5.0, mg./100 ml. in a 1 :10 dilution of serum. Magnesium
turnings
(Mallinckrodt)
dissolved
Standard) dissolved in for calcium. This stock the working standards, 7.5, 10.0, 12.5, and 15.0 in
concentrated
HC1
served as the stock standard for magnesium. This stock solution was diluted to give working standards equivalent to 0.5, 1.0, 2.0, 3.0, and 4.0 mg./100 ml. magnesium in a 1 :25 dilution of serum.
Methods and Results Calcium In Serum
To measure the accuracy with which calcium could he determined iii serum and urine by atomic absorption speetroscopy, both Comparison and recovery studies were carried out. For serum calcium, Willis (2) deproteinized the serum with trichioroacetic acid. He also reported that
206
GIMBLET ET AL.
Clinical
Chemistry
reasonably accurate results could be achieved when serum was simply diluted with either water or 1% (w/v) EDTA. To achieve maximum accuracy with only water as the diluent, he found careful flame adjustment necessary. Since we were seeking the simplest procedure, requiring the least manipulation and the fewest steps, we used 1% EDTA as the diluent. To determine serum calcium, 0.5 ml. of serum is diluted with 4.5 ml. of 1% EDTA. With this diluent, 100.1% of calcium added to serum (5 samples) was recovered. The coefficient of variation (V) was 1.6%; the standard deviation of the sample(s) was 1.62; and s2 was 2.62. These values compare favorably with the 100.1% recovery reported by Willis, with V of 1.9% (calculated by us). To test the relative accuracy of the method further, serum calcium was determined on sera which had been submitted for routine analysis and which constituted a representative sample of abnormal sera; values obtained by atomic absorption were compared with those obtained by oxalate precipitation and permanganate titration as described by ClarkCollip (16) (a modification of the Kramer-Tisdall method (17) for the chemical determination of calcium). If hotil methods measure serum calcium with the same accuracy, tue value obtained by atomic absorption spectrophotometry (y) sllonld equal the value obtained by the oxalate-permanganate method (x). With perfect correlation the best straight line to fit the data would have a slope of 1.0 and an intercept of 0. A series of 56 determinations was run by both methods, on 10 different days. The number of determinations per (lay was or fewer on 7 days, and 8, 9, and 13, respectively, the remaining 3 days. To reduce the amount of calculation, 28 of the parallel sets of determinations (or every other one, beginning with the first), were used for statistical evaluation. Tile straight line wllich best fit these data is described by the equation y = 0.35 + 0.93 x, where y is the value obtained by atomic absorption for serum calcium and x the value by the oxalatepermanganate procedure. The deviation of the slope from 1.0 and the intercept from 0 is not significant, since t is equal to 1.363 and t is equal to 1.267, the critical value of ti% with 25 degrees of freedom being equal to 2.787. The standard deviation s is 0.345. There is no significant difference between these data and those of Willis (.2). These statistical data are presented in Table 1; the statistical parameters on Willis’ data have been calculated by us. The differences between the individual determinations in tile whole set of 56 parallel determinations have been evaluated in another way. It is assumed that the oxalate-permanganate method gives the correct
CALCIUM AND MAGNESIUM
Vol. 13, No. 3, 1967 Table
1. SERUM
CAlciuM
VAI.uEs:
(HEMIcA!,
AXE)
207
A’rOMIc
ABSORPTION
(AAS)
METhODS
Met 1,0(18
limierVo.
V
ept
PRESENT
(‘hemim.
8
1.
lb
0.345
1.565
1.267
2.056
0.016
3.185
INVESTIGATION
AAS-
EDTA
28
0.55
0.953
WILLIS
AAS0
(‘hietmi. AASEDTA
.5
AAS0
*Serurn
22
(2)
0.14
1.006
0.663
0.385’
0.25
1.013
1.788
0.124
2.086
-
deproteinized.
value; then percentage
the value obtained of this value. With
by atomic absorption this approach, the
is calculated values obtained
as a by
atomic absorption average 101.3%, or 1.3% higher than the values obtailled by the oxalate procedure (Table 2). Tile standard deviation of the differences is equal to 4.1%. The value for the significance of the difference between the two methods is 2.210. With 55 degrees of free(tom the critical value of t5% is 2.0002 and of tl% 2.664. There is thus some possibility that by atomic absorptiol1 using 1% EDTA diluent the values for serum calcium are somewhat higher than those obtained by the Clark-Collip method. This difference is of no practical significance, however. In Table 3 a more extensive set of data is shown, comparing serum
Table
2.
CALCIUM
DETERMINATION No.
BY ATOMIC
ABSORPTION,
Recovery
(%)
SERUM
This
study
Willis
5 9 56 5
()
This study Willis (2)
Coi PARED Accuracy
WITH
CHEMIC.\L
METHODS V
CALCIUM
100.1 100.0
-
101.3! 99.6!
-
URINARY
-
1.60 1.93 4.43 5.42
OLCItM
-
This study Willis (4) This study
17
99.6
5.15
Willis (4)
15
100.71
7.45
Percentage tPercentage
4 4
in comparison in
comparison
96.5 100.8
w-itli time oxalate gravimetric with the oxalate-permaaganate
3.96
1.55
method. method.
208 Table
GIMBLET ET AL. 3. SERUM
CALCIUM
BY ATOMIC
ABsORpTION
Clinical
AS COMPARED
WITII
Chemistry
OXALATE-PF.RMANGANATE
PROCEDURE
No.
Serum
1e8t3
(ml.)
38 37
8
Diluent
±3.3 ±3.8
100.1 101.0
4.5 ml., 1% EDTA 4.8 nil., 1% EDTA
0.5 0.2
calcium determinations by atomic absorption with those by the oxalatepermanganate method. The precision with which calcium can be determined, both by atomic absorption using 1% EPTA as the diluent and by tile classical chemical method, is shown in Table 4. V is quite small for replicate determinations performed the same day, and is of the same order for both the instrument procedure and the classical chemical procedure. As might be expected, precision calculated from data collected over several days with a pooled serum as a sample is not as good. The V’s for both methods are approximately 2.5 times as great as those obtained with replicate determinations performed tile same day. A V of this order with the analyses performed on different days represents a quite acceptable degree of precision for this type of analysis. In this regard, it is of illtereSt to compare these coefficients of variation with those obtained for sodium and potassium by flame photometry with the Baird Associates flame photometer (18). These data are shown in Table 5. V for sodium and potassium, as well as for serum calcium by the oxalate-permanganate method, was calculated from data obtained on pooled sera used in our routine quality control program. Since we adopted atomic absorption spectrophotometry as the routine procedure for serum calcium, it has been checked on two different occasions by the classical procedure of oxalate precipitation and perTable
4. CALCIUM
PRECISION
BY AAS
AND
OXALATE-PERMANGANATE
METHODS
Replication Concentration Method
AAS AAS AAS AAS AAS AAS AAS OX-P AAS OX-P
Solution
Standard Standard Standard Standard Standard Stammdard Serum Serum Serum Serum
(mg.I100
2.5 5.0 7.5 10.0 12.5 15.0 9.3 9.6 9.4 9.2
ml-)
Day
No.
V
Same
4 5 5 5 5
1.25 0.35 1.11 0.49 0.42 0.80 0.55 0.63 2.30 2.66
Same Saimie Same Same
Sammie Same Same Different Different
5 6 6 10 10
Vol. 13, No. 3, 1967
Table
5.
CALCIUM
PRECISION
FOR SODIUM
AND
AND
POT.tssluM
Concentration
Sodium Sodiumn Sodium I’otassiumn Potassium Potassium
IN SERUM No.
(mEqIL.)
!,ialysis
on different
BY
FLAME
PHOTOMETRY
of
V
Replicates8 5 9 8 6 8 8
141 139 137 4.6 4.8 4.4
8Amialyzed
209
MAGNESIUM
2.58 1.42 2.63 3.17 1.37 2.27
days.
manganate titration. in both instances essentially the same values for accuracy were obtained as those above: the chemical procedure yielded results equal to 98.4 and 97.3% of those obtained by atomic absorption spectroscopy. In Urine
For the determination of calcium in urine, Willis found lanthanum chloride-a 1% (w/v) solution-to be the best diluent of several tried. This diluent seemed to eliminate the interference from phosphate better than did EDTA or strontium chloride. In preliminary studies with 1% EDTA as the diluent (Table 6), recovery of calcium added to urine appeared better than when 1% lanthanum chloride was used. However, upon more comprehensive testing, dilution with 1% lanthanum chloride (see Table 7) resulted in better agreement with the chemical procedure Table
6.
RECOVERY
OF CALCIUM
ADDED
TO URINE
Recovery Dii uent
EDTA LaCb SrCl2
Table
Test8
(%)
4 4 5
97.4 96.5 90.4
7. ATOMIC
ABSORPTION
V
S
3.62 14.62 31.3
VS. CHEMICAL
1.90 3.82 5.58
METHODS
1.95 3.96 6.18
FOR URINARY
CALCIuM
Ratio .fethods
eompared*
AAS/Ox.G AAS/Ox-G AAS/Ox-P AAS/Ox-P AAS/Ox.P AAS/Ox-P 8Ox-G the Taussky
Diluent
Tests
LaCb
7 7 6 4 7 7
EDTA LaCI3 La18 LaOS3 EDTA
indicates method
time chemical (?O).
oxalate,
(%) 97.9 82.6 96.8 100.8 101.4 83.0 precipitation,
V
14.23
3.78
3.86
31.72 33.42 18.61 30.7
5.62 5.78 4.31 17.5
5.80 5.73 4.25 21.08
-
-
immcimierimtion, gravimetric
-
method;
Ox-P,
210
GIMBLET ET AL.
Clinical
Chemistry
than did dilution with EDTA. Thus our final method utilized a dilutioll with 1% lanthanum chloride. Two different methods were used for the ciieniical determination of urinary calcium. One is the classical macrogravimetric procedure in which calcium is precipitated as tile oxalate, isolated, mcmerated, and finally weighed as the oxide. This procedure was performed for us at another laboratory (19). The other method is that described by Taussky (20), which is essentially an oxalate precipitation-permangallate titratioii procedure applied after the urine has beell extracted with chloroform. In our own evaluation trials of Tanssky ‘s InetilOd, we were able in 7 trials to recover 96.3% of the calcium added to uriiie. rf he coefficient of variation was 3.06. Recovery data and accuracy tests for atomic absorption versus chenlical analyses are reported in Table 2, with comparable data from Willis (4). In Diet and Stool
The diets analyzed were duplicates of those fed to volunteer pregnant patients who were subjects for a metabolic investigation. Tile diets were planned to be adequate for optimum nutrition during pregnancy. The stools were from the same patients. For analysis, diets are first homogenized; samples of tile homogenate are placed in weighed platinum dishes and then w-eighed. They are partially dried at 90#{176}, then dried completely at 130#{176}. The dried material is ashed at 550#{176} in a Thermolyne 2000 furnace. When cool, the ash is dissolved in a few drops of nitric acid and transferred to a 23-mi. volumetric flask containing distilled water; it is then brought to volume with distilled water and appropriate dilutions made. Stools are collected in a No. 5 Lily tub lined with an 8- by 4- by 18-in. polyethylene bag, which weighs about 7 gm. The stool is frozen in this bag, and the frozen stool and bag are weighed together; thus tile weight of the stool is obtained. The frozen stool separates easily and cleanly from the polyethylene bag if it is necessary to i’emove it. In our nietabolic studies, stools from a 48-hr. period are usually combined. To the combined stool as passed, 800 ml. of water is added. Tile siligle plastic bag containing tile frozen combined stools is then encased in two more plastic bags of the same size. The bags are closed by twisting the open ends together, then folding tile twisted ends over. Closure is maintained with a heavy rubber band wrapped around the folded-over ends. This package is placed in a 1-gal. paint can and the cover is closed. Tile can is then shaken for 15 mm. on a paint mixer. A small sample of the re-
Vol. 13, No. 3, 1967
Table
CALCIUM
8. RECOVERY
OP CALCIUM
AND
ADDED
211
MAGNESIUM
TO DIET
AND
STOOL Total
Somnie
(‘a in Saom,mle*
mit. (ym.)
HOMOGENATES
(MEQ)
Ca Recovery
(a
added
Expecied
Found
(% ) t
0.241 0.200 0.245 0.209
100.8 100.5 102.1 96.8
0.463 0.515 0.425 0.438
97.2 100.6 100.2 97.3
DIET HOMOGENATE
5.2658 4.1320
0.184 0.144 0.185 0.161
5.3061 4.6150
0.055 0.055 0.055 0.055
0.239 0.199 0.240 0.216
STOOL HOMOGENATE
5.3295
0.421
5.7858 4.6709 4.9942
0.457 0.369 0.395
0.055 0.055 0.055 0.055
5By
0.476 0.512 0.424 0.450
amialysis, 1 gill. of diet was foumid to 0.079 ,miEq. of Ca. tAverage; for diet, 100,1%; for stool, 98.8%.
comitain
0.034
mmmEq. Ca,
and
1 gum. of
stool
liclliogenate,
sultiiig homogenate is then weighed in a preweighed dish. The remainder of the sample preparation is the same as that t’or diet. Studies of the recovery of calcium added to diet and stool are shown in Table 8. These data indicate the accuracy of the over-all process. For the final dilution, 0.5 ml. of the diluted solution to be analyzed is added to 4.5 ml. of 1% lanthanum chloride. The diets and stools analyzed in tins study are not necessarily representative of all possible diets or stools. Analysis of stools from patients receiving antacids like Amphojel,* which contains aluminum salts, may be subject to error because of tile presence of these salts in the stool. Magnesium
To measure tile accuracy with which magnesium can be determined in serum and urine by atomic absorption spectrophotometry, only recovery experiments, constituting the so-called internal-standard method of evaluation, were employed. In the procedure usually used for serum magnesium, 0.2 ml. of serum was diluted with 3 ml. of diluent. As shown in Table 9, we have found, in agreement with the observations of Willis (3, 4), an enhancement of absorption wilen water is used as a diluent and acceptable recoveries of added magnesium when 0.25% strontium chloride is used as a diluent. Thus tile strontium chloride seems to suppreSS served *Wyeth
tile
enhancenient.
when
15%
Laboratories,
TCA
A similar was used
Philadelphia,
Pa,
enliancenient to precipitate
of absorption the protein.
was This
ob-
effect
212
IMLET Table
9.
RECOVERY
T AL.
OF MAGNESIUM
ADDED
Clinical TO SERUM
AND
Chemistry
URiNE
Recovery Diluent
(%)
Tests
Serum SrCl3 H,O TCA Urine H20 SrCI2
.
101.2 107.8 107.9
2.44 2.74 13.63
1.56 1.65 3.68
1.55 1.53 3.41
5 4
99.4 101.1
8.31 9.17
2.88 3.03
2.90 3.00
OF MAGNESIUM
FROM
DIET
AND STOOL HOMOGENATES Total
Sample
wt.
M9 in
(gm.)
V
4 4 4
10. RECOVERY
Table
5
(MEQ)
Mg
Mg
sampie*
Recovery
added
Expected
(%)t
Found
DIET HOMOGENATE
4.8731
0.058
0.028
0.086
0.083
4.7229
0.057
0.028
0.085
0.083
4.9651
0.060
0.028
0.088
0.087
99
3.6761
0.044
0.028
0.072
0.073
101
103.9 100.0 100.5 106.1 98.1
HOMOGENATE
STOOL
5.4884 5.6433
0.178 0.183
0.027 0.027
0.205 0.210
0.213 0.210
4.9010 4.4241 6.5505
0.159 0.143 0.212
0.027 0.054 0.054
0.186 0.197 0.266
0.187 0.209 0.261
3By 0.0324
analysis, mEq.
Average:
Table
11.
1 gm.
of diet
was
97 98
found
to contain
0.012
mEq.
of Mg,
and
1 gm.
of
stool,
Mg. for diet,
99%;
COEFFICIENTS
for
OF
stool,
VARIATION
Concentration
101.6%.
FOR No.
STANDARDS,
REPLICATED
MAGNESIUM
BY
ATOMIC
ABSORPTION
V
of replicates
ON SAME
DAY
(IG./100
MI..)
0.5
5
1.06
1.0
5
0.56
2.0
5
1.05
3.0 4.0
5 6
0.34 1.64
SERA;REPLICATRD
1.82 1.82 l2
ON DIFFERENT
5 5 16
METHOD
DAYS
(MEQ/L.)
.
1.10 2.07 2.52
Vol. 13, No. 3,
Table
967
CALCIUM
12. RECOVERY
OF
Urine dilution
Mg in urine
213
MAGNESIUM
ADDED TO URINE
MAGNESIUM OR WATER
AND
AS DILUENT Mg added
WITH
(MG./100 Total Mg
STRONTIUM
0.25%
STRONTIUM
CHLORIDE
ML.) Mg found
Recovery
(% I *
CHLORIDE
1 to 100
0.64
0.5
1.14
1.17
103
2 to 100
1.34
0.5
1.84
1.86
101
3 to 100 4 to 100
1.96 2.66
0.5 0.5
2.46 3.16
2.55 3.06
104 97
0.51 0.63 1.23
0.49 0.63 1.23
96 100 100
WATER
0.4 to 100 0.5 to 100 ito 100
0.31 0.38 0.73
0.2 0.25 0.50
2 to 100
1.43
0.50
1.93
1.88
98
3 to 100
2.0
0.20
2.20
2.28
104
5Average:
with
strontium
chloride,
101C% ; with
water,
100%.
observed whether strontium chloride was present or not. The diltient used routinely for serum magnesium determination is 0.23% stroiitium chloride. Recovery of magnesium from diet and stool honiogenates is shown in Table 10. The final solutions used in these analyses consisted of 0.1 ml. of appropriately diluted ash solution in 5.0 ml. of water. The precision with which magnesium can be detei’rnined is shown in Table 11. Again, as expected, the precision obtainable is much better when the replicate analyses are performed on the same day than when they are performed on different days. Table 12 shows the recovery of magnesium from urine with 0.23% strontium chloride and with water used as diluelsts. Water as the diluent appears to give slightly better results statistically, but the difference is not significant. For the sake of simplicity, water is used as the preferred diluent. was
References 1. 2. 3.
4. 5. 6.
Walsh, A., The application of atomic absorption spectra to chemical analysis. Spcclrochim. 4cta 7, 108 (1955). Willis, .J. B., The determimmatioim of metals in blood serum by atomic absorptiomm spectroscopy. I. Calcium. Spectrochim. Acta 16, 259 (1960). Willis, J. B., The determination of metals in blood serum by atomic absorption spectroscopy. II. Magnesium. Spcetroehim. Acta 16, 273 (1960). Willis, J. B., Determination of calcium and magnesium in urine by atomic absorption spectroseopy. Anal. Chem. 33, 556 (1961). Newbrun, E., Application of atomic absorption speetroseopy to time (letermlminatiOn of calciummmin saliva. Nainre 192, 1182 (1961). Dawson, J. B., and Heaton, F. W., The determination of magnesium in biological materials by atomic absorption spectrophotometry. Biochem. J. 80, 99 (1961).
214 7.
8.
9. 10. 11. 12. 13. 14. 15. 16. 17. 18. 19.
GIMBLET fT AL.
Clinical
Chemistry
Herrnmanmm, R., and Lang, W., Analysen vomi Magmiesiumn imi Serum umid im anderen K#{246}rper flussig Keiten mit Hilfe der Absorptions flammen Potometre. Z. Ges. Exp. Med. 135, 569 (1962). Somne biological applications of atomic absorption spectrophotomnetry. Atomic Absorption Newsletter No. 11, March 1963. Horn, D. B., and Latner, A. L., The estimnation of magnesium by atomic absorp’tion spectropliotomnetry. C/in. C/mini. Aeta 8, 974 1963). Stewart, W. K., Hutehmimison, F., and Fleming, L. w., The etimnation of magimesium in Serum alid urine by atomnic speetrophotometmy. J. Lab. C/in. Med. 61, 858 (1963). Decker, C. IF., Aras, A., and Decker, L. B., Determination of magnesium amid calcium iii cerebrospinal fluid by atomic absorption spectroscopy. Anal. Bioc/tein. 8, 344 (1964). Zettmmer, A., and Seligson, D., Application of atomic ahsorptioim spectrophotomnetry in the determimiation of calcium in serum. Gun. Chew. 10, 869 (1964). Leen, M. W., and Atwood, J. G., Design of an atomic absorption speetrophotometer. Pittsburgh Conference Oil Analytical Chemistry, Feb. 27-Mar. 3, 1961. Slavimi, W., A burner-atomizer for atomic absorption speetrophotometry. Atomic Absorption Newsletter No. 10, February 1963. Slavin, W., Sprague, S., and Manning, D. C., The determination of calcium by atomic absorption spectrophotometry. Atomic Absorption. Newsletter No. 15, September 1963. Clark, E. P., and Collip, J. B., A study of the Tisdall method for the determination of blood serum calcium with a suggested modification. J. Biol. Chew. 63, 46i (1925). Kramer, B., and Tisdall, F. IF., A simple technique for the determimiation of calcium and magnesium in small amounts of serum. J. Biol. Ch.em. 47, 475 (1921). White, J. U., Precision of a simple flame photometer. Anal. Chem. 24, 394 (1952). Wolfson, A. H., and Meyer, W., Time detcrniination of calcium iii urine. Personal communication.
20.
Taussky, H. Microchem.
H., Microdeterimiinatioim J. 7, 89 (1963).
and
separatiomi
of
calcium
and
strontium
in urine.