Thin layer chromatography (TLC) [PDF]

Polar compounds will adsorb strongly to the silica and require a polar eluent to move them up the plate, whilst non-pola

3 downloads 6 Views 1MB Size

Recommend Stories


Applied Thin-Layer Chromatography
Love only grows by sharing. You can only have more for yourself by giving it away to others. Brian

2.2.27. thin-layer chromatography
Just as there is no loss of basic energy in the universe, so no thought or action is without its effects,

Pharmacognostical, physicochemical, and high performance thin layer chromatography evaluation
How wonderful it is that nobody need wait a single moment before starting to improve the world. Anne

Recent Advances in Thin-Layer Chromatography of Pesticides
Those who bring sunshine to the lives of others cannot keep it from themselves. J. M. Barrie

Quantitative Analysis of Phospholipids by Thin-Layer Chromatography
Where there is ruin, there is hope for a treasure. Rumi

Analysis of Explosive Mixtures by Thin Layer Chromatography
How wonderful it is that nobody need wait a single moment before starting to improve the world. Anne

Development and Method Validation Densitometry Thin Layer Chromatography for
You often feel tired, not because you've done too much, but because you've done too little of what sparks

Thin-layer cervical cytology
Just as there is no loss of basic energy in the universe, so no thought or action is without its effects,

Thin Cell Layer
Life is not meant to be easy, my child; but take courage: it can be delightful. George Bernard Shaw

Idea Transcript


Thin layer chromatography (TLC)

Chemistry Department

Thin layer chromatography is used extensively for the analysis of mixtures, following the progress of reactions and as a check for purity. The equipment needed to run a tlc is shown below sample

TLC spotter / micropipette beaker with lid

eluent

filter paper

tweezers TLC plate

ruler

pencil

Your sample should contain a small amount of mixture (1-3mg) dissolved in two or three drops of a suitable, volatile solvent (eg. ether or ethyl acetate. Your compound must be soluble in it).

1-3mg of compound dissolved in a few drops of volatile solvent

sample vial

The developing tank for a TLC should consist of a beaker/jar with a lid

watch-glass cover Filter paper should be used saturate the beaker atmosphere with solvent vapour

a clean, dry beaker

The TLC plate consists of a thin layer of silica gel on an aluminium foil backing. Remember to use a pencil for drawing your line. If you use ink it will interfere with your TLC.

Use a pencil and ruler to draw a line about 1 cm from the bottom of the TLC plate.

When setting up the development tank it is a good idea to pour the eluent in before you spot your sample on the TLC plate. This is to ensure that the atmosphere in the beaker is saturated with solvent vapour.

Pour sample in

eluent

After you have added the eluent to the beaker, swirl it around the edges and replace the watch-glass cover

The eluent to be used depends on the sample. Polar compounds will adsorb strongly to the silica and require a polar eluent to move them up the plate, whilst non-polar compounds will move up the plate with a relatively non-polar eluent

If the spots are close to the eluent front then decrease the eluent polarity.

Least polar

Aim to get the spots moving about half way up the plate for best results.

Light petroleum (b.pt. 40-60°C) Cyclohexane Pentane Toluene Dichloromethane Diethyl ether Ethyl acetate (Ethyl ethanoate) Acetone (propanone) Ethanoic acid (Acetic acid)

Most polar

Methanol

If the spots remain on or near the baseline increase the eluent polarity.

When spotting you sample onto your TLC plate, you should label each spot in pencil. Below the labels ‘s’, ‘m’ and ‘wk’ are used to represent strong, medium and weak samples, but you are free to label them however you wish.

Label the spots

Spot your samples onto the TLC plate using a micropipette

TLC plate with all three sample spots

Tweezers should be used to transfer the TLC plate into the developing tank. Allow the back of the plate to rest on the sides of the beaker at a slight angle.

Make sure the level of eluent is below the base line on the TLC plate

Replace the watch-glass cover and leave the plate to develop.

Allow the solvent front to rise up the TLC plate in a horizontal line until it reaches about 1 cm from the top of the plate.

Remove the TLC plate from the developing tank and mark the solvent front on the TLC plate before it starts to evaporate.

solvent front

Unless the material being analysed is coloured the plate will need to be treated in order to see the spots and measure the distance travelled. UV light can be used to observe the spots and for some analysis the plate will need to be stained.

TLC plate under UV light. Observation of the additional spots.

TLC plate in normal light. Observation of the original spots.

TLC plate with the original and additional spots circled.

The retention factor (Rf) is a useful measurement which can be made from a developed TLC plate. It is the relation between the distance the compound has travelled and the distance travelled by the solvent front.

Distance travelled by the compound

Distance travelled by the solvent