Using EDTA to Ameliorate the 'Prozone Effect' [PDF]

Human complement factors, notably C1, C3, C4, interfere with the Luminex assay often resulting in falsely low results. (

4 downloads 3 Views 336KB Size

Recommend Stories


Edta EDTA
You often feel tired, not because you've done too much, but because you've done too little of what sparks

Investigation on the Ability of Sulfonolipid from Salinibacter ruber to Ameliorate the Toxic Effect of
In the end only three things matter: how much you loved, how gently you lived, and how gracefully you

(PDF) The Compound Effect
Be who you needed when you were younger. Anonymous

Final Programme - ERA-EDTA 2018 [PDF]
May 25, 2018 - Liu, Inst Nephrology, Southeast Univ, Renal Dept, Nan Jing City, China. ...... Chang Seong Kim, Hong Sang Choi, Ha Yeon Kim, Eun Hui Bae, Seong ...... Seung Duk Hwang, Byung Chul Yu, Moo Yong Park, Soo Jeong Choi, Jin ...... A STABLE G

[PDF] The Gaslight Effect
Stop acting so small. You are the universe in ecstatic motion. Rumi

[PDF] The Compound Effect
Do not seek to follow in the footsteps of the wise. Seek what they sought. Matsuo Basho

[PDF] The Compound Effect
Knock, And He'll open the door. Vanish, And He'll make you shine like the sun. Fall, And He'll raise

Tris Acetato EDTA
If your life's work can be accomplished in your lifetime, you're not thinking big enough. Wes Jacks

(EDTA) chelating agent (minimum)
I cannot do all the good that the world needs, but the world needs all the good that I can do. Jana

ERA-EDTA Registry
Ego says, "Once everything falls into place, I'll feel peace." Spirit says "Find your peace, and then

Idea Transcript


Using EDTA to Ameliorate the ‘Prozone Effect’ in UK NEQAS for H&I’s ‘HLA Antibody Specificity Analysis’ Scheme Felicity May, Tracey Rees, Deborah Singleton & Christopher Darke

Introduction

Serum Treatment Methods Used

UK NEQAS for H&I’s ‘HLA Antibody Specificity Analysis’ scheme (Scheme 3) assesses participants’ ability to define HLA specificities in antisera using the same method(s) they employ for clinical samples. 81 laboratories participated in scheme 3 in 2015, and all used a Luminex-based method. The majority of participants used Luminex alone (n = 65), but some used Luminex in combination with other methods (Luminex + CDC n = 14; Luminex + ELISA n = 1; Luminex + ELISA + Flow Cytometry n = 1). Human complement factors, notably C1, C3, C4, interfere with the Luminex assay often resulting in falsely low results (the so-called prozone effect). A number of test modifications can be used to overcome this problem, including EDTA treatment of sera, heat inactivation, and DTT treatment. These methods are increasingly widely used, particularly in highly sensitised patients.

Participant Survey A survey of serum treatment techniques was circulated to Scheme 3 participants. The survey comprised of 10 questions relating to serum treatment techniques with particular focus on EDTA treatment methodology.

Results 42 of the 81 participants (51.9%) returned responses to the survey.  61.9% of laboratories (n = 26) used some form of serum treatment  38.1% of laboratories (n = 16) currently have no routine treatment measures in place The following methods were used by participants; some laboratories applied several methods depending on the clinical sample:  EDTA addition (EDTA) - 73.1%, n=19/26  Serum dilution (Dil) - 38.5%, n=10/26  Heat inactivation (HI) - 15.4%, n=4/26  DTT addition (DTT) - 11.5% n=3/26

HI only EDTA & DTT Dil, HI & DTT Dil & HI EDTA & HI Dil only EDTA & Dil EDTA only None

1 1 1 1 1 4 4 13 16 0

5

10

15

20

Number of Partipants Concentrations of EDTA stock solutions used included 6%, 0.1M, 0.2M and 0.5M. EDTA to serum ratios ranged between 1:10 and 1:20. Of the 19 laboratories that used EDTA, 84.2% (n=16) added it directly to the sera while 3 laboratories added it to the wash buffer. Of the 16 labs that added EDTA to serum, 11 (68.8%) tested samples immediately after treatment while 5 routinely stored sera prior to testing (3 at -20°C and 2 at 4°C). 5 laboratories commented that some samples were stored routinely whilst others were tested immediately depending on, e.g. sample urgency. Of the 11 laboratories that immediately tested samples after EDTA treatment, 8 (72.7%) applied a minimum EDTA/serum incubation time of 5 to 15 minutes. Three laboratories had no formal serum EDTA treatment time.

Comment Importantly, no laboratories that employed a treatment method (i.e. EDTA, dilution, heat inactivation or DTT) had unsatisfactory Scheme 3 results for 2015.

Further Information Full information on all UK NEQAS for H&I schemes is available at www.neqashandi.org.uk or contact the Scheme Manager at [email protected]